China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
  1. China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
  2. China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
  3. China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
  4. China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale
  5. China Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators for sale

Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators

Price Negotiable
MOQ 1
Delivery Time 3-5work days
Model Number SGX-VOX
Payment Terms L/C, D/A, D/P, T/T, Western Union, MoneyGram

Product Details

Product Specification

Payment Terms L/C, D/A, D/P, T/T, Western Union, MoneyGram Zero current (offset) <0.1% O2
Delivery Time 3-5work days Linearity Linear
Minimum Order Quantity 1 T90 response time < 5 seconds
Output signal 9-13mV in Air* Model Number SGX-VOX
T99.5 response time <50 sec Measurement Range 0-100% O2

Product Description

Gas Oxygen Sensor SGX-VOX Electrochemical Gas Sensor 0-100% Oxygen (O2) Sensor With On Board Temperature Compensation For Ventilators

 

CROSS - SENSITIVITY:

DATA Toxic gases at TLV levels will have no cross-sensitivity effect on SGX-VOX oxygen sensors. At very high levels (i.e. % levels), highly oxidising gases (e.g. ozone, chlorine) will interfere to the extent of their oxygen equivalent, but most other commonly occurring gases will have no effect.

 

ACID GASES :

Acid gases such as CO2 and SO2 will be absorbed by the electrolyte and tend to increase the flux of oxygen to the electrode. This gives an enhanced oxygen signal of approximately 0.3% of signal per 1% CO2. SGX-VOX sensor is not suitable for continuous operation in concentrations of CO2 above 25%. The SGX-VOX sensor is not designed for use in applications where anaesthetic gases are present.

 

Temperature Range -20°C to +50°C
Pressure Range 500 to 2000 mbar
Operating Humidity Range 0% to 99% RH (non condensing)
Long Term Output Drift <5% per annum
Recommended Storage Temp 0°C to 20°C